====== Implementation of continuous cal at Ke ====== | ====== Implementation of continuous cal at Ke ====== | ||
- | edit /usr2/control/skedf.ctl | + | In ''station.prc'', take out calls to ''caltsys_man'', ''systemp12'' |
- | <code> | + | In ''setupsx'', set <code>cont_calq=on</code>. In ''caltsys'', remove ''caltsys_man'' and replace with ''tpicd=tsys''. Remove ''ifagc'' and ''bbd_gain=all,agc'' calls |
- | cont_cal ON | + | |
- | + | ||
- | + | ||
- | </code> | + | |
- | + | ||
- | in station.prc, take out calls to caltsys_man, systemp12 | + | |
- | + | ||
- | in setups, set <code>cont_cal=on</code>. In caltsys, remove ''caltsys_man'' and replace with 'tpicd=tsys''. Remove ifagc and bbd_gain=all,agc calls | + | |
+ | In ''point.prc'', in ''initp'', ''cont_cal=on'' | ||
- | in point.prc, in initp, ''cont_cal=on'' | + | ''/usr2/control/skedf.ctl'' added <code>tpicd NO 300</code> and changed to ''cont_cal ON''. Tsys will be logged every 30 sec by default. |
- | edited both files ''/usr2/control/rxg_files'' and changed creation date, in 7th line+, commented out previous tcal value and inserted measured values from Jamie. Measured Tcal values (as of April 15 2014): | + | Edited both files in ''/usr2/control/rxg_files'' and changed creation date, in 7th line+, commented out previous tcal value and inserted measured values from Jamie. Measured Tcal values (as of April 15 2014): |
<code> | <code> | ||
RCP Cont-CALs | RCP Cont-CALs | ||
Line 27: | Line 20: | ||
X6U 8857 7.3 0.4 | X6U 8857 7.3 0.4 | ||
X7U 8916 7.7 0.4 | X7U 8916 7.7 0.4 | ||
- | X8U 8929 7.3 0.5 | + | X8U 8937 7.3 0.5 |
- | X8L 8937 7.3 0.5 | + | X8L 8929 7.3 0.5 |
S1U 2230 15.5 0.3 | S1U 2230 15.5 0.3 | ||
S2U 2250 11.0 0.3 | S2U 2250 11.0 0.3 |