 +===Cryo===
  
 +Warmed and pressure drop
 +
 +- commence vaccuming S/X using turbo pump
 +
 +- turn off the SX cryodyne in the diesel room so that the blue motor stops
 +
 +- turn off the L-band cryodyne by pulling out the lead to the L-band
 +cold head in the focus cabin
 +
 +- use hastings vac gauge and Edwards oil filled vac pump to pump L-band
 +dewar while it warms(dear valve open whilst pumping).
 +
 +- top up the helium supply to 300 psi if the compressor is running OK.
 +If has shut down, increase the pressure slightly then restart the
 +compressor and increase the pressure to 300 psi supply.
 +
 +- locate the leak
 +
 +- power down the compressor
 +
 +- replace leaking hose
 +
 +- power up the compressor
 +
 +-top helium up to 300 psi
 +
 +- commence s/x cooling by turning on cryo S/X power in diesel room
 +
 +- start L-band cryodyne
 +
 +- 3 hours of cooling with vac pump on L-band is required
 +
 +- close l-band dewar valve
 +
 +- remove Edwards pump.
 +
 +-close SX dewar valve and turn off turbo