====== RF and IF Configuration ======
Check the Status Monitor to see if the Cross-over switches and RCP IF Attenuators are at the default levels:
* At all sites, cross-over switches should be **Direct**
* IF Attenuator settings should be as follows
* At Hobart:
* S RCP : 0.0 dB
* X RCP : 10.0 dB
* ASCII code if setting manually: ''@@TT''
* At Katherine
* S RCP : 15.0 dB
* X RCP : 8.5 dB
* ASCII code if setting manually: ''__QQ''
* At Yarragadee
* S RCP : 10.0 dB (changed to 15.0 dB to keep power levels - Lucia, Jun 2015)
* X RCP : 7.0 dB
* ASCII code if setting manually: ''TTNN''
Check the crossover switch and attenuator settings via the [[operations:starting_sysmon|System Monitor GUI]].
The main thing is that the target power levels (listed in the next session [[operations:documentation.dbbcsetupivs_v2|DBBC setup]]) are OK. If values aren't as above, but power levels are attained without radical attenuation, then all is well. Otherwise, the attenuation level should be adjusted using the ''rfpic'' and ''ifpic'' commands in the Field System. Documentation for these commands is here:
* [[operations:documentation:pcfs_rfpic|rfpic documentation]]
* [[operations:documentation:pcfs_ifpic|ifpic documentation]]
but each station has setup procedures for the default settings. Cross-over switched to Direct:
xodirect
IF attenuation to nominal:
ifnom
Occasionally these commands may report a communications error. If this happens, try re-issuing the command.
If the issue persists, it may be necessary to restart Monica's Rxmon and IFBox on pcfs.
* [[restart_monica_rxmon_ifbox|Restart Rxmon and IFBox]]
To set these values manually, use the nc (netcat) command and appropriate ASCII code given above. For example, to set the attenuators at Yarragadee, log in as //oper// to //pcfsyg// and type:
nc pcfsyg 60002
S
Prompt will return: 'values', so type in:
TTNN
Then type "Ctrl-C" to quit