17:55 UT Experiment started late due to issues with startup and inexperience with setting up the 26m (first setup). First valid scan was 139-1803b , missed 5.1 gb of data (Arwin).
17:55 UT ERROR ch -302 v1 frequency does not check with requested value. This error only appeared once. (Arwin)
17:55 UT ERROR ch -308 ve total power integrator overflow. Consistently getting this error. (Arwin)
19:08 UT Data recorded is on par with the summary file. (Arwin)
00:58:47 UT (DOY 140) Wind stowed, schedule halted (Warren)
03:01:57 UT recording started, schedule resumed after wind stow (Warren)
07:54 UT in checklist disk_pos off by approx 29gb probably due to late start and earlier wind stow (Liz)
09:51 UT “ERROR ch -309 v6 alarm” is on occurred 8 times between 09:45:14 and 09:47:45. Could only find one other occurrence of this in the wiki from RD1411 in 2014. At that time it was stopped by “valarm” but it seems to have gone away on its own at this point. Doesn't seem like something anyone was particularly concerned about however, just in case, only scan affected would be 140-0946.(Liz)
11:00 UT “ERROR ch -309 v6 alarm is on” error x4 from 11:00:44 to 11:01:47 UT. Scans affected part of 140-1101b.(Liz)
12:07 UT “ERROR ch -309 v6 alarm is on” again x12 between 12:01:43 to 12:05:42. It seems to occur in periods. (Liz)