Apply a tone of 764
MHz at -15 dBm to one input 1 of all four CoMos
Use dbbc_config_file_104_calibration.txt
Set dbbcifa/b/c/d=3,agc,1 (i.e. input 3, filter 1)
Once gains have stabilised, turn off the agc loop with dbbc=dbbcgain=all,man
Run “calibration” in DBBC Control
dbbc=dbbcgain=all,agc
The results are written to C:\ as yyyy-ddddThhmmss-calib.txt
The file has 5 columns: phase, CoMo A/B/C/D value. Pick the phase at the null past the first peak.