Sometimes the DBBC needs reprogramming because channels have been dropped or spectra look strange. To do this, preferable in a gap between scans, follow the following steps:
Applications --> Auscope Hobar --> VNC -->dbbchb
DBBC2 Control DDC v104_2.exe
. Close it and then restart it from the desktop. When prompted to reconfigure press “y”, it takes about 2 minutes.dbbc=pps_sync
You should then see a row of LEDs on the DBBC flash in sync, once a second.
fmset
In the interface, type S
followed by y
, then press the ESC key.
mk5=dot?
and check that you see syncerr_eq_0
and FHG_on
(or FHG_off
). If you see something like syncerr_gt_3
then fmset
may need to be run again.
maserdelay
and clkoff
commands which read the Agilent counters.ERROR sp -4 GPIB Device time-out on response c2
To fix this, try re-setting the counters. In the oprin
window, type
counter
and try the clkoff
and maserdelay
commands again.