Line 23: Line 23:
  
   - mk5ke: bank B back plane failure. Affected disks 4/5 and 6/7. Do not use for conditioning.   - mk5ke: bank B back plane failure. Affected disks 4/5 and 6/7. Do not use for conditioning.
 +  - mk5-2ke: module doesn'​t go all the way in to bank B.
   - DBBC: "stuck bit" and bit statistics problems related to VSI OUT connectors: ch15 (S) with VSI1 (rear) -- was problematic from the beginning, ch5 (X) with VSI2 (front) emerged after the inspection and refitting of FILA OUT board.   - DBBC: "stuck bit" and bit statistics problems related to VSI OUT connectors: ch15 (S) with VSI1 (rear) -- was problematic from the beginning, ch5 (X) with VSI2 (front) emerged after the inspection and refitting of FILA OUT board.
    
  - DBBC: suppresses the amplitude of 10-kHz pcal tone (first) in each band.  - DBBC: suppresses the amplitude of 10-kHz pcal tone (first) in each band.
- 
- - Modules: hob+025/​6000 has fault in disks 6/7; bkg-0015/​1480 took forever to condition. Are labelled as broken and prepared for repair. bkg-0015/​1480 should be reprofiled into higher capacity disk. 
  
  - Maser: replaced the '​switches'​ unit and AC/DC converter, cavity tuning and frequency adjustment performed, new standby batteries were installed. ​  - Maser: replaced the '​switches'​ unit and AC/DC converter, cavity tuning and frequency adjustment performed, new standby batteries were installed. ​
 +
 +**Status:**
 +
 + - mk5-ke is a main recorder connected to VSI2 on FILA OUT. Channel 5 have a stuck bit. 
 + - 5MHz signal for pcal is connected directly to the second output of the maser through the synch port.
 +
 {{tag> backend}} {{tag> backend}}
  
Line 58: Line 63:
  
  - Control room: mk5 light for reading module labels is no longer connected to exterior light switch on ipske. It now switches on ''​pdu1ke.phys.utas.edu.au''​ and is labelled ''​light int''​  - Control room: mk5 light for reading module labels is no longer connected to exterior light switch on ipske. It now switches on ''​pdu1ke.phys.utas.edu.au''​ and is labelled ''​light int''​
 +
 + - Installed 2 of power distribution units (PDU). These are like an 8 way power board that have individually switched outputs. They can be switched from a web interface at the following 2 links:
 +\\ [[http://​pdu1ke.phys.utas.edu.au/​outlet.htm|pdu1ke.phys.utas.edu.au/​outlet.htm]]
 +\\ [[http://​pdu2ke.phys.utas.edu.au/​outlet.htm|pdu2ke.phys.utas.edu.au/​outlet.htm]]
 +
 +The user is admin and the password is the same one we use for the internet power switches.
 +\\ PDU1KE so far has wifi, cduke (network maint computer) and the interior light.
 +\\ PDU2KE so far has mk5-2ke and the GPIB controller.
  
  
/home/www/auscope/opswiki/data/attic/blog/blogkatherine_maintanance_trip_june_15-24.1403520641.txt.gz · Last modified: 2014/06/23 10:50 by liza