 The temperature and humidity sensors in the Katherine control and maser rooms have been recalibrated using the data from the loggers (24/7/14, ~1030 UT). Data from Monica recorded before this time can be rescaled using the formulae below: The temperature and humidity sensors in the Katherine control and maser rooms have been recalibrated using the data from the loggers (24/7/14, ~1030 UT). Data from Monica recorded before this time can be rescaled using the formulae below:
  
-Control Room:+**Control Room:** 
 TrueRH = MonicaRH*1.2 + 4,7 TrueRH = MonicaRH*1.2 + 4,7
 +
 TrueTemp=MonicaTemp*0.9 + 0.66 TrueTemp=MonicaTemp*0.9 + 0.66
  
-Maser Room:+**Maser Room:** 
 TrueRH = MonicaRH*0.83+32 TrueRH = MonicaRH*0.83+32
 +
 TrueTemp=MonicaTemp TrueTemp=MonicaTemp
 +
 +The temperature and humidity sensors in the Katherine pedestal and hub have been recalibrated using the data from the loggers (29/7/14 ~0140 UT). Data from Monica recorded before this time can be rescaled using the formulae below:
 +
 +**Pedestal:​**
 +
 +TrueRH = MonicaRH*0.95 + 23
 +
 +TrueTemp=MonicaTemp*0.92 + 0.32
 +
 +**Hub:**
 +
 +TrueRH = MonicaRH * 0.83  +27
 +
 +TrueTemp=MonicaTemp *0.93 +0.77
 +
 +**Feed**
 +
 +Feed temp & RH sensors still not working.
  