AOV035

Hobart 26m

Disk VSN: HOB+0125/32000/1024 | Data volume at beginning: 15.9 TB

  • 1725UT - Cryo monitoring system not working at Ho (ERROR ma -4 Device rx timed-out on response). I have replaced the cryomon and rxmon procedures with nops in EXPERIMENT procedure library for the time being, because otherwise it waits for each rx measurement to time out. I have also commented out the rxmon command in the midob procedure in STATION library, as that is also waiting for timeout, and screwing up scan timing. (Patrick)
  • 1730UT - Experiment started OK (Patrick)
  • 1800UT - Windstows (fill in!) (Patrick
  • 1925UT - Field system time out of sync (ERROR sc -13 setcl: formatter to FS time difference 0.5 seconds or greater). Fixed by running sy=run setcl offset in oprin.

Katherine 12m

Disk VSN:HOB+0015 | Data volume at beginning: 0.0 GB

  • 1730UT - No system temperature measurements. Restarted DBBC pc, which fixed it. (Patrick)
  • 1730UT - Experiment started OK (Patrick)

Yarragadee 12m

Disk VSN: HOB+0131 | Data volume at beginning: 0.0 GB

  • 1730UT - Experiment started OK (Patrick)
  • 1730UT - no connection to Yg maser (Patrick)