Line 7: Line 7:
   * 17:00 UT - Experiment started OK. (Tiege)   * 17:00 UT - Experiment started OK. (Tiege)
   * 18:47 UT - Alarm due to delay through DBBC changing, will monitor to see if the delay stabilises. (Tiege)   * 18:47 UT - Alarm due to delay through DBBC changing, will monitor to see if the delay stabilises. (Tiege)
-  * 18:50 UT - Delay immediated ​stabilised at ~0.4 us, with variation from 0.398 us through to 0.403 us. Going to leave it be as long as it does not start drifting again. (Tiege) +  * 18:50 UT - Delay immediately ​stabilised at ~0.4 us, with variation from 0.398 us through to 0.403 us. Going to leave it be as long as it does not start drifting again. (Tiege) 
-  * 19:26 UT - delay difference drifted again to 0.435 us, going to try and see if restarting DBBC software helps. ​halted ​schedule during 256-1925a, first scan after reset will be 256-1944. (Tiege)+  * 19:26 UT - delay difference drifted again to 0.435 us, going to try and see if restarting DBBC software helps, seems to be staying at the nominal ~0.257 us post-resetHalted ​schedule during 256-1925a, first scan after reset will be 256-1944. (Tiege) 
 +  * 22:59 UT - Have noticed that in the time between the previous two checklists ( time where I was asleep) the delay difference has very slowly drifted from ~0.26 us to 0.285 us, gradual enough that is hasn't yet set off the alarm. Jim will be taking over soon so I will mention it to him. (Tiege) 
 +  * 23:25 UT: DBBC was still causing a drift so rebooted the machine. Jim 
 +  * 23:37 UT: Restarted with scan 256-2347. OK now. Jim
  
 ** Katherine ** ** Katherine **
Line 20: Line 23:
   * 18:25 UT - Total missed/​dodgy scans list: 256-1700b to 256-1733b inclusive were recorded onto USN-0069 using mk5ke (potentially dodgy), scans 256-1742 through to 256-1802 were missed while switching to mk5-2ke, 256-1808 and 256-1809 WERE RECORDED on the new disk (USN-0017), then 256-1814 through to 256-1827b were missed attempting to correct clkoff issue. (Tiege)   * 18:25 UT - Total missed/​dodgy scans list: 256-1700b to 256-1733b inclusive were recorded onto USN-0069 using mk5ke (potentially dodgy), scans 256-1742 through to 256-1802 were missed while switching to mk5-2ke, 256-1808 and 256-1809 WERE RECORDED on the new disk (USN-0017), then 256-1814 through to 256-1827b were missed attempting to correct clkoff issue. (Tiege)
   * 18:30 UT - Experiment should be running as normal from  scan 256-1833b onwards. Just note that 256-1808 and 256-1809 were both recorded to this module also. (Tiege)   * 18:30 UT - Experiment should be running as normal from  scan 256-1833b onwards. Just note that 256-1808 and 256-1809 were both recorded to this module also. (Tiege)
 +  * 00:58 UT - Observations stopped from 00:55:48 to 00:58:10 to put in another module in bank A. Jim
 +
 +  * **0720UT to 1030UT** (JS)
 +     * MK5-2KE crashed as well, this was a UPS issure but in the meantime MK5 was fixed
 +     * GPIB fixed, clkoff/​maserdelay now working
 +     * Module USN-0017 transferred from mk5-2ke to mk5 bank A, recording will continue on this
 +     * mk5 and dbbc reset, and synced with fmset
 +     * First new scan 257-1039. Missed scans 257-0720b through 257-1035c (215GB deficit) ​
  
  