User Tools

Site Tools


hardware:dbbc3

This wiki is not maintained! Do not use this when setting up AuScope experiments!

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
hardware:dbbc3 [2019/02/19 02:53]
Guifre
hardware:dbbc3 [2019/06/18 00:39] (current)
Guifre [Check calibration's quality]
Line 6: Line 6:
   * VNC connection with standard pwd   * VNC connection with standard pwd
   * remote desktop with Windows XP user/pwd   * remote desktop with Windows XP user/pwd
 +=== Installation DDC_V_v123
 +Notes provided in 20.11.2018
 +
 +  - Copy **executable** file 'DBBC3 Control_V_v123.exe'​ into bin folder or Desktop
 +  - Copy **firmware** files  '​dbbc3_ddc-V_v123*.bit'​ into DBBC_CONF/​FilesDBBC
 +  - Copy **config** files: ​
 +      * '​dbbc3_config_file_V_123.txt'​
 +      * '​config_ddc_V.txt'​ / '​config_ddc_V_12bbc.txt' ​
 +      * '​config_adb3l.txt'​
 +      * '​ddc_v_core3H_#​.fila10g'​ / ddc_V_12bbc_core3H_#​.fila10g'​
  
 ==== Start DBBC3 Control ==== ==== Start DBBC3 Control ====
Line 16: Line 26:
  
 Launch DBBC Client v4. exe to communicate and launch the dBBC3 commands. Launch DBBC Client v4. exe to communicate and launch the dBBC3 commands.
 +
 +Keep editing the config_adb3l.txt file with the new values obtained.
  
 ==== Calibration of the DBBC3 ==== ==== Calibration of the DBBC3 ====
Line 23: Line 35:
   - Calibrate gain   - Calibrate gain
  
-===== Calibrate offset =====+=== Calibrate offset === 
 + 
 + 
 +  dbbcif{a,​b,​c,​d,​e,​f}=2,​60 ​    //- Set a strong attenuation for each of the IF// 
 +  dbbcif{a,​b,​c,​d,​e,​f} ​         //- Check that each TPI value is in the order between 1000 to 8000// 
 +  cal_offset={1,​2,​3,​4,​5,​6} ​    //- Launch the cal_offset program for each the Core boards// 
 + 
 +One offset calibration per board, so in total it is needed to conduct the offset calibration 6 times. 
 + 
 +=== Calibrate ​gain === 
 + 
 +  dbbcif{a,​b,​c,​d,​e,​f}=2,​agc,​1,​34000 ​    //- Set an intermediate TPI value (34000) and wait until value reach// 
 +  dbbcif{a,​b,​c,​d,​e,​f}=2,​man ​        //- Set dbbcif with agc off // 
 +  cal_gain={1,​2,​3,​4,​5,​6} ​    //- Launch the cal_gain program for each the Core boards// 
 + 
 +Again 6 calibrations are needed.  
 + 
 +=== Calibrate delay === 
 + 
 +Do it after the gain and offset ​calibration. Use similar initial settings for each board as in the Calibration gain procedure. 
 + 
 +  dbbcif{a,​b,​c,​d,​e,​f}=2,​agc,​1,​34000 ​     
 +  dbbcif{a,​b,​c,​d,​e,​f}=2,​man 
 +  cal_delay={1,​2,​3,​4,​5,​6} ​    //- Launch the cal_delay program for each the Core boards// 
 + 
 +==== Check calibration'​s quality ==== 
 + 
 +  enablecal=on,on,on 
 +  enableloop 
 +  core3hstats=board_nr
  
-<WRAP center round tip 60%> +<WRAP center round important 50%> 
-  * dbbcif{a,​b,​c,​d,​e,​f}=2,60     //- Set a strong attenuation for each of the IF// +  * All sample powers in a 2% variation 
-  * dbbcif{a,​b,​c,​d,​e,​f} ​         //Check that each TPI value is in the order between 1000 to 8000// +  * All bitstats following ​proportion of 18-32-32-18% 
-  * cal_offset={1,​2,​3,​4,​5,​6} ​    //Launch ​the cal_offset program for each the Core boards//+  * Auto-correlation in the order of 2 Million units.
 </​WRAP>​ </​WRAP>​
  
-Proceed 6 times with the offset ​calibration.+Last calibration ​made by Guifré on Hb on 19.02.2019.
  
 +Last calibration made by Guifré on Yg unit on 18.06.2019.
 +- board1 = offset OK, gain KO,
 +- board2 = offset OK, gain OK, delay KO
 +- board3 = offset OK, gain OK, delay KO
 +- board4 = offset OK, 
 +- board5
 +- board6 ​
/home/www/auscope/opswiki/data/attic/hardware/dbbc3.1550544790.txt.gz · Last modified: 2019/02/19 02:53 by Guifre