 +===Webcams===
 +The webcams are useful as a visual check that the antennas are out of park, moving and slewing, when live pages, or remote interfaces fail.
 +There is a wireless webcam at each Auscope site; the username and password are the usual for observers,
 +http://​webcamho.phys.utas.edu.au
 +http://​webcamhb.phys.utas.edu.au
 +http://​webcamke.phys.utas.edu.au
 +http://​webcamyg.phys.utas.edu.au
  
 +==== 26 m Telescope ====
 +==Receiver Position==
 +In Otter interface [r] resets box in focus cabin. If receiver position display is "​zero"​ indicates error (all bits zero) indicative of reset required.
 +Above fault also means auto receiver position not working.
  
 +===Cryo System===
 +
 +Action plan after warmed and severe pressure drop. Compressor has cut out as pressure went below 150 psi
 +
 +- commence vaccuming S/X using turbo pump
 +
 +- turn off the SX cryodyne in the diesel room so that the blue motor stops
 +
 +- turn off the L-band cryodyne by pulling out the lead to the L-band
 +cold head in the focus cabin
 +
 +- use hastings vac gauge and Edwards oil filled vac pump to pump L-band
 +dewar while it warms(dear valve open whilst pumping).
 +
 +- top up the helium supply to 300 psi if the compressor is running OK.
 +If has shut down, increase the pressure slightly then restart the
 +compressor and increase the pressure to 300 psi supply.
 +
 +- locate the leak. Leak detector goes yellow for heavy molecules, red for light gas. Keep skin away from tip, and don't breathe near tip.
 +
 +- power down the compressor
 +
 +- replace leaking hose
 +
 +- power up the compressor, must have > 150 psi to start
 +
 +-top helium up to 300 psi once compressor running
 +
 +- commence s/x cooling by turning on cryo S/X power in diesel room
 +
 +- start L-band cryodyne
 +
 +- 3 hours of cooling with vac pump on L-band is required
 +
 +- close l-band dewar valve
 +
 +- remove Edwards pump.
 +
 +-close SX dewar valve and turn off turbo