This wiki is not maintained! Do not use this when setting up AuScope experiments!
This is an old revision of the document!
Chair: Ross
Attended:
Apologies:
Absent:
General Issues
Yg said they don't get information about the disk space required for an experiment from the new schedule format (.vex experiments?).
Yg also said that sometimes the module spreadsheet is not updated (w.r.t. which modules are assigned to experiments).
Handover Issues
r4848 (ke) Can't access maserke vnc, but can ping the machine.
crds96 (ke) received “Warning: error tc -301 continuous cal not enabled for dbbc” for each scan. Fixed continuous cal not enabled error by changing cont_cal=off to cont_cal=on in the setup01 function of the procedure file, based off the previous crds95 procedure file.
aov024 (ho) Was getting “ERROR sc -13 setcl: formatter to FS time difference 0.5 seconds or greater ”. Ran sy=run setcl offset into the newsmerd fs window (oprin)
r1847 (ke) noticed coordinates had not changed, checked to see the FS had crashed. Restarted FS and the schdule, missing data from 0729 UT till 08:22 UT. It appears log monitor had also stopped responding.
r1847 (yg) Humidity sensor appears to not be working.
/home/www/auscope/opswiki/data/attic/operations/agenda2018_06_28.1530074638.txt.gz · Last modified: 2018/06/27 04:43 by Ross Turner