This shows you the differences between two versions of the page.
Both sides previous revision Previous revision Next revision | Previous revision | ||
operations:documentation.dbbcsetupivs [2014/03/24 23:53] Vasaant Krishnan |
operations:documentation.dbbcsetupivs [2016/05/16 12:48] (current) Bryn Emptage |
||
---|---|---|---|
Line 33: | Line 33: | ||
You now need to set up the IF system. The commands to do this depend on the IVS observation and the station. | You now need to set up the IF system. The commands to do this depend on the IVS observation and the station. | ||
- | For R1, R4, CRF, AUSTRAL and RDV experiments on they should be: | + | For R1, R4, CRF, CRDS, AUSTRAL and RDV experiments on they should be: |
<code> | <code> | ||
dbbcifa=1,agc,1 | dbbcifa=1,agc,1 | ||
Line 43: | Line 43: | ||
- | For APSG, CRDS, OHIG and T2 experiments, for all stations they should be: | + | For APSG, OHIG and T2 experiments, for all stations they should be: |
<code> | <code> | ||
dbbcifa=1,agc,1 | dbbcifa=1,agc,1 | ||
Line 52: | Line 52: | ||
</code> | </code> | ||
- | The target values for the total power readings is 56000. this can be checked by the command: ifa, ifb, ifc or ifd. If the levels are too high or low, then adjust the attenuation on the IF box. Note that the levels on IF C (e-remote control command: ifc) are a bit low. Don't worry about it. | + | Recommended DBBC power levels can be found [[hardware:dbbc.targetifpower|here]]. Current target values for the total power readings is 56000. this can be checked by the command: ifa, ifb, ifc or ifd. If the levels are too high or low, then adjust the attenuation on the IF box. Note that the levels on IF C (e-remote control command: ifc) are a bit low. Don't worry about it. |