The temperature and humidity sensors in the Katherine control and maser rooms have been recalibrated using the data from the loggers (24/7/14, ~1030 UT). Data from Monica recorded before this time can be rescaled using the formulae below:

Control Room:

TrueRH = MonicaRH*1.2 + 4,7

TrueTemp=MonicaTemp*0.9 + 0.66

Maser Room:

TrueRH = MonicaRH*0.83+32

TrueTemp=MonicaTemp