R1740

Hobart 12m:

17:00 UT: Experiment Started OK. Disk position 1799.86 GB at start. (Lucas)

17:00:10 UT 2016.138.17:00:10 WARNING: No PCFS activity for 0.1 minutes.

2016.138.17:00:10 Check VNC session to see if Field System software is running

2016.138.17:00:10 Check network connection.

Restarted field system from IP, also monica, reconfigure dbbc amd log monitor/e-control.

This lead to a new error: (e.g)

2016.138.21:52:35.28?ERROR qk -211 Tsys value for device eu overflowed or was less than zero.

2016.138.21:52:39.19?ERROR ib -10 ibcon initialization failed, gpib devices inaccessible.

Tried to restart GRIB from IP however this did not work.

Jim managed to restart/reconfigure everything; include re adjusting the attenuators which did not restore after fs reboot (is that correct Jim)?

First 'good' scan: 138-1943 755 2052-474 216 84 CW 19:43:06 19:43:49 0:43 300.0 *

All scans before missed. (Lucas)

  • 23:50 UT: Manually power cycled GPIB to try and fix error ib -10 - Talk light was constantly on (possible communication jam). After power cycle - talk light is no longer constantly on. Error persists. (Bryn)
  • 02:06 UT: Restarted fs to try fix error ib -10. Missed scans 139-0212 and 139-0217. Unsuccessful - Will reboot pcfs during next gap in schedule. (Bryn)
  • 02:37 UT: Rebooted pcfs. (Bryn)
  • 02:50 UT: error ib -10 fixed, errors with dbbc communication - cycling power. (Bryn)
You could leave a comment if you were logged in.
/home/www/auscope/opswiki/data/attic/handover/r1740.1463540360.txt.gz · Last modified: 2016/05/18 02:59 by Warren Hankey