DBBC-HB
Replaced CPU unit and Xilinx programmer 29/9/10. Now able to communicate with FPGAs and apparently configure them. Currently exits with “Register_Card error=-13”. All cards are securely inserted into the backplane. Replugging does not seen to affect the result - possible backplane fault?
DBBC-AUSCOPE2
Timing issues - clock-1024.exe was not starting correctly when DBBC Control was run and this lead to serious timing errors from the PPS, where a drift of ~0.5 microsec/sec was seen in the PPSMon-GPS difference. Fixed temporarily by restarting clock-1024.exe from C:\DBBC\bin. The problem has not recurred in such a rapid formbut the PPSmon-GPS output is often drastically incorrect. We have measured an offset of 100-200 microseconds where the maser PPS-GPS is 12 microseconds. pps_sync does not correct this. The problem can sometimes be resolved by power cycling the electronics and reconfiguring. At the best, a slow trend is seen in the PPSMon-GPS, of a few microseconds in tens of minutes. This is still too high for use in geodesy.
Missing channels - Some channels (9,13) appear to be permanently stuck with the recorded data consisting entirely of -1 (1-bit recording). The problem is not fixed by power cycling the electronics. Other channels intermittently appear distorted and full of weak, spurious tones (1-8).
LEDs on front panel show that they occasionally go into a fault state without apparent cause. No PPS is seen in the FPGA module and the device requires a power cycle of the electronics to function. The only LED lit up on the fault modules is the 10th from the bottom (10 kHz retuning enable?)
dbbcifc always reads slightly low (52000-54000). We have switched CoMos without make any difference and amplifying the signal does not affect this.
This DBBC is currently powered on with 10 MHz & 1 PPS applied but no input IFs.
CoMos do not seem to have good isolation on the input selector switch. Inputs without any attached IFs can still read as having significant total power.
DBBC-AUSCOPE3
Currently in use for IVS testing observations. Performance seems ok except that channels 3 and 4 intermittently drop out, with recorded data consisting solely of -1 (1-bit recording). There are no indications of this error state in the LEDs of the front panel. NOTE: as of 4 Nov 2010, not always channels 3 and 4. This time channels 1, 9, 11 and 12 dropped out:
dbbcifd power always reads slightly low (52000-54000). We have switched CoMos without make any difference and amplifying the signal does not affect this.
This machine is currently in use (4th October-6th October). Do not make any changes to the configuration until after this IVS run!