Sometimes the DBBC needs reprogramming because channels have been dropped or spectra look strange. To do this, preferable in a gap between scans, follow the following steps:
Applications --> Auscope Hobar --> VNC -->dbbchb
DBBC2 Control DDC v104_2.exe. Close it and then restart it from the desktop. When prompted to reconfigure press “y”, it takes about 2 minutes.dbbc=pps_sync
You should then see a row of LEDs on the DBBC flash in sync, once a second.
fmset
In the interface, type S followed by y, then press the ESC key.
mk5=dot?
and check that you see syncerr_eq_0 and FHG_on (or FHG_off). If you see something like syncerr_gt_3 then fmset may need to be run again.
maserdelay and clkoff commands which read the Agilent counters.ERROR sp -4 GPIB Device time-out on response c2
To fix this, try re-setting the counters. In the oprin window, type
counter
and try the clkoff and maserdelay commands again.