Chair: Katie

Attended:

Apologies: Prad

Absent:

General Issues

Handover Issues

r1867 : Katherine getting large delay - but can be fixed using using the 'counter' command or resetting the dbbc.

ohg114 : Hobart

  • Unordered List Itemifc and ifd were not on default values which lead to a 'error di -101 no default for if input' when 'setupsx' was entered during the setup. Fixed by copying values from previous ohg summary file to the current one.
  • Receiver cryogenic pressure high, vacuum pump started
  • No Tsys readings, reason unclear (JMc)

r1868 : (Katherine) alarms going off for illegal state error and unable to read systemclock1. antenna=open antenna=operate fixed the issue

ohg116: Katherine - Large difference between formatter and maserdelay.