This shows you the differences between two versions of the page.
| Both sides previous revision Previous revision Next revision | Previous revision | ||
|
operations:agenda2019_05_16 [2019/05/16 01:24] Lim Chin Chuan |
operations:agenda2019_05_16 [2019/05/16 03:14] (current) Lim Chin Chuan |
||
|---|---|---|---|
| Line 1: | Line 1: | ||
| **Chair:** Lim | **Chair:** Lim | ||
| - | **Attended:** | + | **Attended:** Momtaz Simin, Mas, Tiege, Patrick, Jamie, Katie |
| - | **Apologies:** Ross | + | **Apologies:** Ross, Earl, Jay, Lucas |
| **Absent:** | **Absent:** | ||
| Line 12: | Line 12: | ||
| * Please remember to create handover note page during setup. | * Please remember to create handover note page during setup. | ||
| * No connection to Yg and Hb maser. | * No connection to Yg and Hb maser. | ||
| + | * Go to pcfs terminate the e remote control and restart if cannot connect. | ||
| + | * End message, if problem exists, please mention/describe the error. | ||
| **Handover Issues** | **Handover Issues** | ||
| Line 23: | Line 25: | ||
| CRD102: | CRD102: | ||
| * Ho: 1307UT - fmset puts the field system time 6.2 seconds ahead of the Mark5 and Computer time, this seems to result in scans ending (and starting) ~5 seconds ahead of schedule (when going off the log time). We are in the last 50 mins of the experiment so I am going to leave as is rather than spend 25 mins restarting machines and end up losing more data than would be (potentially) saved. (Tiege) | * Ho: 1307UT - fmset puts the field system time 6.2 seconds ahead of the Mark5 and Computer time, this seems to result in scans ending (and starting) ~5 seconds ahead of schedule (when going off the log time). We are in the last 50 mins of the experiment so I am going to leave as is rather than spend 25 mins restarting machines and end up losing more data than would be (potentially) saved. (Tiege) | ||
| - | * | + | * on going issue. Jamie will solve it tomorrow. Get the 26m to log the time difference |
| R1894: | R1894: | ||
| * Yg: pcfsyg disk filled up 15 minutes before experiment started. Needed to remove some old epi/png files from the auto_corr_test directory. (Tiege) | * Yg: pcfsyg disk filled up 15 minutes before experiment started. Needed to remove some old epi/png files from the auto_corr_test directory. (Tiege) | ||
| - | * | + | * Cronjob to clean up directory everyday |
| + | * New machine coming in a few months with new system. | ||
| + | * Might still need to use the old system. | ||
| AOV035: | AOV035: | ||
| Line 33: | Line 37: | ||
| * Fixed by resetting TTY distributor with “reset” switch on old S/X MarkIV rack. Serial communications from Hobart to s/x receiver are still patched through this rack. (Warren and Brett) | * Fixed by resetting TTY distributor with “reset” switch on old S/X MarkIV rack. Serial communications from Hobart to s/x receiver are still patched through this rack. (Warren and Brett) | ||
| * 1725UT - wx command not working. (Patrick) | * 1725UT - wx command not working. (Patrick) | ||
| + | * Path failure. | ||
| + | * Ups failed (lost power) | ||
| + | * Working now. | ||